Fluke 50-Series II (Model 51) Thermocouple Thermometer with Single Input

Fluke  |   Mfr# FLUKE-51-2 60HZ  |   Item# DO-91500-04

$299.99 USD/EACH

Are you a reseller or buying in bulk? Call us today.


  • Rugged design with splash- and dust-resistant housing
  • Readout in °F, °C, or Kelvin (K)
  • Meters are ruggedly designed with a splash- and dust-resistant housing.
  • Sleep mode increases battery life
Dual backlit display provides readout of channel temperature, differential temperature (dual-input models), and minimum/maximum/average readings with time stamp.


Available for sale in the United States and Canada only.

What's Included:
one type K bead thermocouple per input channel and protective holster.
Product Type Thermocouple meter
Max temperature (º C) 1372
Temp range Type J: -346 to 2192°F (-210 to 1200°C)
Type K: -328 to 2501°F (-200 to 1372°C)
Type T: -418 to 752°F (-250 to 400°C)
Type E: -238 to 1832°F (-150 to 1000°C)
Min temperature (º F) -418
Max temperature (º F) 2501
Min temperature (º C) -250
Accuracy Above -148°F (-100°C) J, K, T, E, and N: ±0.05% of rdg plus 0.5°F (0.3°C)
Dimensions 3.375 in W x 6.875 in H x 1.5 in D
Display LCD
Resolution K <1000°: 0.1°F (°C) K 1000°: 1°F (°C)
Model FLUKE-51-2 60HZ
Qty/ea 1
Manufacturer number FLUKE-51-2 60HZ
Brand Fluke
Power Three AA batteries
Battery life 1000 hours
Make TEMPERATURE

Accessories
Qty/Price
Availability

at $53.25 $47.93 USD/EACH

In stock.

at $9.25 USD/PACKAGE of 4

In stock.

at $44.99 USD/EACH

In stock.

at $39.99 $35.99 USD/EACH

In stock.
Calibrations
Qty/Price
Availability

at $135.00 USD/EACH
Usually ships in 9 days.

at $168.00 USD/EACH
Usually ships in 7 days.




Technical Resources
 
Service & Support
Still have questions
Contact Our Experts
for FREE product
application selection
and support.
or call 1-800-358-5525
Custom Ordering Solutions
Custom Ordering Solutions
Let us find the exact product you need. We have access to suppliers full lines - for products beyond what you see here. Call 1-800-358-5525
Questions and Answers

Do you have questions about this product?

Get answers from real customers and in-house experts with AnswerBox.

7 Questions | 7 Answers
Displaying questions 1-7
  • Q:

    Is the probe that comes with the product ok use in small volumes of fluid?
    Asked on 1/22/2013 by Brad

    1 answer

    • STAFF EXPERT

      A:

      The probe must be used in PFA-compatible environments and is *not* suitable for liquid immersion.
      The included probe is a Type K Bead Thermocouple designed for general purpose temperature measuring applications. It has an extremely fast response time, due to the low thermal mass of the exposed tip. The 40-inch (1-meter) PFA cable is terminated with a Type K miniature thermocouple connector with 0.312-inch (0.792-mm) pin spacing. The probe can be used with any temperature measuring instrument that is designed to accept Type K thermocouples and that has a miniature connector input. Above 260°C (500°F), PFA will emit toxic fumes.
      If you need to submerge in fluids, please see probe part number 08466-82 instead.

      Answered on 1/22/2013 by Jen from Technical Support
  • Q:

    Is there an A/C adapter available for this instrument?
    Asked on 1/9/2013 by AMACK from St. Louis, MO

    1 answer

    • STAFF EXPERT

      A:

      No, an AC adapter is not available for this unit, it does not have a port for one either.

      Answered on 1/11/2013 by Jen from Technical Support
  • Q:

    We have purchased the FLUKE Model 51 Series II. How do we go about calibrating thermocouples J Type using this instrument?
    Asked on 1/9/2013 by Tuan from Jessup, MD

    1 answer

    • STAFF EXPERT

      A:

      Calibration consists of two parts: 1) Calibration of the A/D converter; 2) Reference junction calibration, using the ice bath test set.A/D Calibration.

      To calibrate the A/D portion of the thermometer, follow these steps:
      1) Connect the calibrator to the T1 input of the thermometer using copper (nonthermocouple)
      wire and a copper thermocouple connector.
      2) Turn the thermometer on, with a small, blunt probe, press the calibration button
      located on the back and in the center of the thermometer.
      3) The menu on the display changes to “CAL A-d”.
      4) Press 'up' or 'down' to select different menu options.
      5) With CAL A-d selected, press 'enter'. The thermometer prompts “80”. Set the
      calibrator to +80.000 mV and allow the output to stabilize.
      6) Press 'enter'. Wait while it takes A/D readings.
      7) When the thermometer prompts for “-8”, set the Calibrator to -8.0000 mV. Allow the
      output to stabilize.
      8) Press 'enter' on the thermometer and wait while it takes A/D readings. The
      thermometer now shows the measured value in the bottom portion of the display.
      Millivolts appears on the left side and microvolts appears on the right side.
      9) If the measurement is from –7 995.0 to –8 005.0, press 'enter' and the meter will
      return to the calibration menu and display CAL rEF. If the measurement is outside –
      7995.0 to –8005.0, press 'enter' once to return to the calibration menu. Using
      'up' or 'down', select CAL A-d and try the A/D calibration procedure a second time
      starting at step 5.
      10) Using 'up' or 'down', select “CAL donE”, then press 'enter' to store the calibration in
      memory.

      To calibrate the reference junction, follow these steps:
      1) Prepare an ice bath.
      2) Connect the equipment to T1 using K-type thermocouple wire
      and a K-type thermocouple connector. Allow the setup to thermally stabilize at least
      5 minutes after inserting the thermocouple connector into the thermometer.
      3) Turn the thermometer on and select the thermocouple type using the 'setup', 'up', 'down',
      and 'enter' buttons.
      4) Enter calibration mode (refer to the “A/D Calibration” section, step 2).
      5) Using 'up' or 'down', select CAL rEF and press 'enter'.
      6) Press 'up' or 'down' until the upper display matches the known temperature.

      Answered on 1/10/2013 by James from Technical Support
  • Q:

    When 1st purchased, does the fluke 51-2 60HZ come already calibrated and certified?
    Asked on 2/15/2011 by Anonymous from Sacramento, Ca

    1 answer

    • STAFF EXPERT

      A:

      The calibration of every unit is check before it leaves the factory; however, if you require an NIST-traceable certificate, it must be purhcased separately. This calibration with certificate is available as our 17002-10.

      Answered on 2/17/2011 by Josh from Technical Support
  • Q:

    How do I get info on calibrating the Fluke-51 when needed?
    Asked on 11/28/2010 by Anonymous

    1 answer

    • STAFF EXPERT

      A:

      Fluke recommends sending the unit back to them for calibration on a yearly basis. We also provide NIST certificates of calibration from our Metrology Lab.

      Answered on 12/1/2010 by Mike from Tech Support
  • Q:

    When you buy the unit does a probe come with it -- can you please provide me with information on the probe type available for this unit.
    Asked on 9/30/2009 by Anonymous

    1 answer

    • STAFF EXPERT

      A:

      This unit comes with a type K bead thermocouple probe. For another possible probe please see part number 08516-55. This is a general-purpose type K thermocouple probe. This thermometer will accept type J, K, T, & E thermocouple probes with a miniconnector. Please contact our Application Specialists at the number indicated on our website if you have additional questions about selecting the appropriate probe for your application.

      Answered on 9/30/2009 by Jaime from Technical Support
  • Q:

    Does this unit have recording capabilities?
    Can information be uploaded to a computer to be printed?
    Asked on 8/24/2009 by Anonymous

    1 answer

    • STAFF EXPERT

      A:

      This unit does not have recording capabilities, please use part number 91500-08 for the datalogging unit with software and cable package 91500-81 for computer connection.

      Answered on 8/26/2009 by Jen from Technical Support