NIST-Traceable Calibration; Air/Gas Flowmeter

Item# DO-17080-00

$262.00 USD/EACH

Are you a reseller or buying in bulk? Call us today.


Usually ships in 11 days.
Qty

  • Testing range 0.1 to 80,000 sccm at uncertainties as low as (K=2) ±0.2% of reading
  • Testing includes measurement at five to seven points — depending on the meter — equidistant across the range of the flowmeter
  • Certificate includes measured data, graphs, and correlations (if required) to meet the most stringent quality and scientific requirements
Measuring flow is important to process control, with flowmeters used in applications ranging from chemical line monitoring to mining to the oil and gas industry. At InnoCal®, our computerized, automated standards allow us to perform NIST-traceable mass and correlated flow measurements using air or nitrogen gas. When correlation to other gases is needed, our metrologists will provide you with correlated data or a gas correlation table. Other specifics on flowmeter calibrations:
  • Testing range  0-500 LPM air flow
  • Testing includes measurement at 5 to 10 points—depending on the meter—equidistant across the range of the flowmeter
Typical brands that we calibrate include Cole-Parmer®, Alicat, Aalborg Instruments, Gilmont®, Sierra® Instruments, Bios International, and many more.

To meet your calibration needs:

  • All calibration certificates contain as-found and as-left data, estimated measurement uncertainties (emu’s), and test uncertainty ratios (TUR’s) at no additional cost!
  • This NIST-traceable calibration allows you to confirm that the specific asset you purchase meets the manufacturer’s published specification
  • EPA, USDA, ISO, and GLP/GMPs standards require regular NIST-Traceable calibration of all test equipment being used in critical applications, FDA compliant
  • Receive your product calibrated and ready to use rather than sending it to a third-party or internal calibration lab and incurring additional delays
All measurements reported are traceable to SI units via national standards maintained by NIST and performed in compliance with ANSI Z540, ISO17025, and MIL-STD-45662A. InnoCal is accredited to ISO/IEC 17025 by the American Association for Laboratory Accreditation (A2LA) under certificate number 1746.01. This accreditation is your assurance that the calibration is performed in a qualified, controlled laboratory.
Product Type Flow Calibration
Pipe size 3/16" to 1/4" ID
Qty/ea 1
Brand InnoCal

Accessories
Qty/Price
Availability

at $62.00 USD/EACH

Usually ships in 12 days.




Technical Resources
 
Service & Support
Still have questions
Contact Our Experts
for FREE product
application selection
and support.
or call 1-800-358-5525
Custom Ordering Solutions
Custom Ordering Solutions
Let us find the exact product you need. We have access to suppliers full lines - for products beyond what you see here. Call 1-800-358-5525
Questions and Answers

Do you have questions about this product?

Get answers from real customers and in-house experts with AnswerBox.

4 Questions | 5 Answers
Displaying questions 1-4
  • Q:

    How long is the standard calibration period?
    Asked on 4/26/2016 by QC Tim from Philadelphia, PA

    1 answer

    • STAFF EXPERT

      A:

      The standard calibration period is 12 months or one calendar year from the date of calibration. You can specify a different interval to meet your needs if you process or documentation protocol requires.

      Answered on 4/27/2016 by Jose from Technical Support
  • Q:

    My air flowmeters are on a 3-year cal cycle. Can Cole Parmer accommodate that on the calibration certificate?
    Asked on 9/26/2013 by Rick from US

    1 answer

    • STAFF EXPERT

      A:

      We can indicate a custom recalibration cycle on the NIST-cert.

      Answered on 9/26/2013 by James from Technical Support
  • Q:

    I have a 0 - 140 lpm acrylic flow meter for air but I need to use it for helium, can you tell me the correlation between reading and the actual flow rate for helium?
    Asked on 5/6/2011 by Wil from Kingman, Arizona

    2 answers

    • Staff Reviewer

      A:

      We would only be able to look into correlation data if we knew the part number of the meter itself. Different factories have different correlation equations/formulas to determine this. Please contact us by phone or email for further assistance.

      Answered on 5/9/2011 by Jen from Technical Support
    • A:

      Please refer to:
      http://www.abzar-control.com/fa/technical-information/

      Answered on 8/6/2013 by Farzad from Iran
  • Q:

    Do you have a carbon monoxide flow chart for this EW-17080-00 flow meter? If so, what is the range of the flowmeter?
    Asked on 3/4/2011 by Anonymous from Lancaster, PA

    1 answer

    • Staff Reviewer

      A:

      This is a service part number, not a flowmeter. This service is to calibrate a flowmeter. If you need assistance in choosing an appropriate flow meter please call or email us with your specifications.

      Answered on 3/7/2011 by Jen from Technical Support
Displaying questions 1-4